• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC XOMED INC. XPS® BUR; BUR, EAR, NOSE AND THROAT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC XOMED INC. XPS® BUR; BUR, EAR, NOSE AND THROAT Back to Search Results
Model Number 1883672HS
Device Problem Material Fragmentation (1261)
Patient Problem Unspecified Tissue Injury (4559)
Event Date 07/08/2022
Event Type  Injury  
Event Description
It was reported that during procedure during drilling, the tip of bur broke in several parts and got sucked into the frontal sinus and end stayed stuck into the shaver.Broken pieces of the reported product remain inside the frontal sinus and retrieved.A second incision was performed.Used a second motor and a second bur to remove the fragments.The motor was broken too.The procedure was completed with backup product(s).Patient alive with injury.There was surgical delay of 1 hour.
 
Manufacturer Narrative
Medical safety review: medical safety review task states that the reported event and its severity are known and labeled, per ifu 68e3858 f, which states excessive pressure applied to bur may cause bur fracture.Should a bur fracture occur during use; extreme care must be exercised to ensure that all fragments of the bur are retrieved and removed from the patient.Unremoved bur fragments may cause tissue damage to the patient.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XPS® BUR
Type of Device
BUR, EAR, NOSE AND THROAT
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer (Section G)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer Contact
glen belmer
6743 southpoint drive north
jacksonville, FL 32216
6122713209
MDR Report Key15201094
MDR Text Key297635725
Report Number1045254-2022-00461
Device Sequence Number1
Product Code EQJ
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 08/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1883672HS
Device Catalogue Number1883672HS
Device Lot Number02233663589
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/19/2022
Initial Date FDA Received08/10/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-