Catalog Number ROB00079 |
Device Problem
Application Program Problem: Parameter Calculation Error (1449)
|
Patient Problems
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/21/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that during a navio assisted tka surgery in the tibia free collection stage, mapping does not work.When he started collecting, mapping green points were disappeared and then yellow tibia bone, not gray, appeared as if it had already been collected.It is unknown how the procedure was completed.The patient outcome is unknown.
|
|
Manufacturer Narrative
|
Internal complaint reference (b)(4).
|
|
Manufacturer Narrative
|
H3, h6: the navio surgical system korea, part number rob00079, serial number (b)(6)., used for treatment was not returned for evaluation.A relationship between the reported event and the device could not be established.A complaint history review for similar reported/confirmed complaints has identified prior events.A review of manufacturing records indicate the device met all specifications upon release into distribution.The failure mode and associated risk have been anticipated within the risk file and the documented risk level is still within the product risk profile.Although the reported problem was not confirmed through a visual or functional evaluation, factors contributing to the reported symptom may have been associated with a software bug.Based on the investigation, the need for a corrective action is not recommended or required at this time.Should any additional information be received the complaint will be reopened.The failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
|
|
Search Alerts/Recalls
|
|