Model Number PM3562
Device Problem Failure to Interrogate (1332)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/29/2022
Event Type  malfunction  
Event Description
It was reported that the device was unable to communicate with radiofrequency (rf) telemetry.Abbott technical support alleged that the event was due to telemetry lockout.An attempt was made to reestablish rf telemetry by performing a software download, but it was not successful.The patient was in stable condition and will continue to be monitored.
 
Manufacturer Narrative
The reported event of no radiofrequency (rf) telemetry communication using the merlin transmitter was confirmed.Based on the provided information, it was determined that the device was in rf lockout, per design specification.The software that was redownloaded was an older version, which is why the lockout could not be resolved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ALLURE MP CRT-P
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key15204439
MDR Text Key304925101
Report Number2017865-2022-17320
Device Sequence Number1
Product Code NKE
UDI-Device Identifier05414734510097
UDI-Public05414734510097
Combination Product (y/n)N
Reporter Country CodeEN
PMA/PMN Number
P030035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 12/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2023
Device Model NumberPM3562
Device Catalogue NumberPM3562
Device Lot NumberP000146378
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/29/2022
Initial Date FDA Received08/10/2022
Supplement Dates Manufacturer Received12/03/2022
Supplement Dates FDA Received12/19/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/30/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-