Upon inspection of the um-s20-20r-3, ultrasonic probe, performed by olympus personnel, it was determined the probe tip was length was found to be out of specification.There was no patient involvement associated with the reported event.
|
This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation.The subject device was inspected and no visible damaged was noted.The ultrasonic transducer was undamaged, and the ultrasonic medium did not leak from any holes in the tip or insertion sheaths.The probe tip was 1.5 mm longer than the specification.The reported problem was therefore confirmed.A review of the device history record (dhr) found no deviations that could have caused or contributed this issue.The dhr confirmed that the subject device was shipped in compliance with the tip clearance standards.The tip clearance may have been extended because the blade (flexible shaft) was shifted to the end of the base by a physical load (e.G.Bending).The following were identified as potential contributory factors; (a) the device may have been stored in a bent state such that the diameter was 20 cm or less, (b) during reprocessing, the inner blade may have dropped making a return to the normal state impossible, or (c) temperature or humidity effects may have induced the deformity.Ultimately, the investigation presumed that the tip clearance (the transparent sheath potion of the probe tip) may have been stretched.Although not definitive, this was established as the most likely root cause of the reported problem.Olympus will continue to monitor field performance for this device.
|