• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CORPORATION IMUFLEX BLOOD BAG SYSTEM; IMUFLEX WB-SP BLOOD BAG SYSTEM WITH LR FILTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CORPORATION IMUFLEX BLOOD BAG SYSTEM; IMUFLEX WB-SP BLOOD BAG SYSTEM WITH LR FILTER Back to Search Results
Catalog Number 1BBLGQ506A6
Device Problems High Test Results (2457); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/11/2022
Event Type  malfunction  
Event Description
The customer reported an elevated white blood cell (wbc) content in the whole blood product.There was not a transfusion recipient or patient involved at the time of the unit processing, therefore no patient information is reasonably known at the time of the event.The device is not available for return because it was discarded by the customer.
 
Manufacturer Narrative
Investigation is in process.A follow up report will be provided.
 
Event Description
The customer reported an elevated white blood cell (wbc) content in the whole blood product.There was not a transfusion recipient or patient involved at the time of the unit processing, therefore no patient information is reasonably known at the time of the event.The device is not available for return because it was discarded by the customer.
 
Manufacturer Narrative
This report is being filed to provide additional information in h.6 and h.10.Investigation: we have not received the set for investigation; therefore, we conducted the following investigations based on the information provided.In making the blood bags concerned, sealed bags are filled with solution and the line is assembled.These bags are sterilized, stacked, and placed into the blister packs.The top film of each blister pack is heat-sealed.In making leukoreduction filters, filter membranes are formed and put in filter housings.We reviewed the manufacturing record of the lot number in question.There was no equipment trouble causing the issue concerned and we confirmed that the equipment had operated properly and no anomalies had been observed.We also reviewed the manufacturing records regarding particulate removal rates of filter membranes that were likely to be related to filtration performance.It was confirmed that all filter membranes conformed to all standards.Release testing, which includes measurements of solution concentration and volume and a visual inspection, is performed on the product concerned on a sample basis.We reviewed each testing and inspection record of the production number and confirmed that there were no anomalies in all release testing items.The product conformed to the standards.Root cause: from the above-mentioned investigation results, we did not observe any abnormalities in the manufacturing process of the lot number concerned.The filter membranes of the product concerned consist of a rough pore pre-membrane (the first filter membrane) and minute pore main-membranes (the second through eighth filter membranes).According to the data at the time of designing of the product in question, there is a possibility of white blood cell contamination when the particulate removal rates are low.However, as mentioned above, the particulate removal rates of the lot number concerned were within the standards and did not show a low tendency.Therefore, we were not able to identify the cause of the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMUFLEX BLOOD BAG SYSTEM
Type of Device
IMUFLEX WB-SP BLOOD BAG SYSTEM WITH LR FILTER
Manufacturer (Section D)
TERUMO CORPORATION
fujinomiya 418-0 004
JA  418-0004
Manufacturer (Section G)
TERUMO CORPORATION, FUJINOMIYA FACTORY OF TERUMO CORP
818 misonodaira
fujinomiya 418-0 004
JA   418-0004
Manufacturer Contact
makoto yoshikawa
818 misonodaira
fujinomiya 418-0-004
JA   418-0004
MDR Report Key15235339
MDR Text Key305508828
Report Number9681839-2022-00055
Device Sequence Number1
Product Code CAK
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/01/2023
Device Catalogue Number1BBLGQ506A6
Device Lot Number211217AF
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/22/2022
Initial Date FDA Received08/16/2022
Supplement Dates Manufacturer Received08/24/2022
Supplement Dates FDA Received09/09/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-