• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ENDURITY PACEMAKER, DR, MRI RESTRICTED; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ENDURITY PACEMAKER, DR, MRI RESTRICTED; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM2162
Device Problem Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/27/2020
Event Type  malfunction  
Event Description
Related manufacturer reference number: 2017865-2020-23013.During an in clinic follow-up, high pacing impedance and a loss of capture were observed on the right atrial (ra) lead.Abbott technical support was contacted and suspected the electrical anomalies may be due to inadequate connection between the ra lead and the header of the pacemaker.No intervention was performed at this time.The patient was stable and will continue to be monitored.
 
Manufacturer Narrative
This mdr report product is for an ous event identified as same/similar during a retrospective review as a result of abbott¿s investigation.
 
Manufacturer Narrative
Correction: g3 - date received by manufacturer should have been aug 18, 2022, not dec 1, 2020, as reported in follow-up report number 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURITY PACEMAKER, DR, MRI RESTRICTED
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key15237382
MDR Text Key302930855
Report Number2017865-2022-19525
Device Sequence Number1
Product Code DXY
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P880086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 08/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2022
Device Model NumberPM2162
Device Lot NumberP000107910
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/01/2020
Initial Date FDA Received08/16/2022
Supplement Dates Manufacturer Received12/01/2020
08/24/2022
Supplement Dates FDA Received08/18/2022
08/24/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/12/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-