Model Number UHI-4
Device Problems Display or Visual Feedback Problem (1184); Incorrect, Inadequate or Imprecise Result or Readings (1535); Failure to Auto Stop (2938); No Flow (2991)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/18/2022
Event Type  malfunction  
Manufacturer Narrative
Upon inspection and testing of the returned device, the reported issue (device cannot stop automatically) was confirmed and found to be caused by a cr board (printed circuit board) failure.It was observed the dampers were broken due to which internal sockets of the cr board were deformed, which resulted in disappearing of the front panel display, and failure of the device to insufflate.It was observed the gas volume could not be measured accurately due to a broken manifold unit.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available, this report will be supplemented accordingly.
 
Event Description
It was reported the subject device would not stop automatically.The device is an olympus loaner.There was no patient or procedure involvement.The subject device was returned to an olympus service center for evaluation.During inspection and testing, it was observed the front panel display disappeared, the subject device would not insufflate, and the gas volume could not be measured accurately.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Please see the updates in sections h4, h6, and h10.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 6 years since the subject device was manufactured.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely the device does not switch to automatic stop, the front panel display disappeared, and the device does not feed air is due to a broken socket of the printed circuit board due to a damaged dumper.It is likely that the device does not measure the amount of air feeding is due to a broken manifold unit.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama,
odakura, nishigo-mura,
nishishirakawa-gun,, fukushima 961-8 061
JA  961-8061
Manufacturer Contact
masaharu hirose
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8-061
JA   961-8061
426422891
MDR Report Key15237729
MDR Text Key304896352
Report Number3002808148-2022-01078
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170324147
UDI-Public04953170324147
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/20/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/18/2022
Initial Date FDA Received08/16/2022
Supplement Dates Manufacturer Received08/29/2022
Supplement Dates FDA Received09/19/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/08/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-