• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL BRK¿ TRANSSEPTAL NEEDLE, 71 CM LENGTH; TROCAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL BRK¿ TRANSSEPTAL NEEDLE, 71 CM LENGTH; TROCAR Back to Search Results
Model Number 407200
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/11/2022
Event Type  malfunction  
Event Description
During the atrial fibrillation procedure, the needle was noted to be fractured upon removal from the patient.The transseptal needle had been used for an atrial septal puncture, but the puncture was not successful.The transseptal needle was exchanged to complete the procedure.
 
Manufacturer Narrative
One brk transseptal needle was received for evaluation.The stylet and tip were also returned.The brk needle had been fractured at its distal end.Additional information revealed the needle had been reshaped.The brk needle instructions for use state ¿do not alter this device in any way¿.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the needle tip fracture is consistent with not following the instructions for use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRK¿ TRANSSEPTAL NEEDLE, 71 CM LENGTH
Type of Device
TROCAR
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key15239887
MDR Text Key302958740
Report Number3008452825-2022-00556
Device Sequence Number1
Product Code DRC
UDI-Device Identifier05414734205092
UDI-Public05414734205092
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K072278
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number407200
Device Catalogue Number407200
Device Lot Number8334378
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/11/2022
Initial Date FDA Received08/17/2022
Supplement Dates Manufacturer Received09/30/2022
Supplement Dates FDA Received10/11/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/10/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-