SMITH & NEPHEW ORTHOPAEDICS LTD UNKN BIRMINGHAM HIP RESURFACING (BHR) CUP (UNKN TYPE); PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL Back to Search Results
Device Problem Biocompatibility (2886)
Patient Problem Metal Related Pathology (4530)
Event Date 07/11/2022
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event Description
It was reported that after a bhr resurfacing construct had been implanted on (b)(6) 2010, the patient experienced metallosis with a big pseudo-tumor.This adverse event was treated by a revision surgery on (b)(6) 2022, in which the bhr femoral head was explanted.Further information is unknown at the time.
 
Manufacturer Narrative
H3, h6: it was reported that a hip revision surgery was performed due to metallosis with a big pseudotumor.As of today, the implanted devices, all of which were used in treatment have not been returned for evaluation.As no device part and batch numbers were provided for investigation, a complaint history review, manufacturing record review, device labelling / ifu, or escalation actions review could not be performed.If more information is received, this investigation will be reopened.A risk management review was performed for the acetabular cup.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.No further actions are required at this time.No relevant supporting clinical information has been provided to assist with a clinical investigation.Without the requested clinical information, a thorough medical investigation cannot be rendered.It should be noted that the implant remained in situ for 12 years.It cannot be concluded that the reported issue was caused by a mal function of the components.The patient's current condition is unknown.With the information provided, the impact to the patient beyond the surgery and recovery cannot be determined.Should any additional clinical information be provided, this complaint will be re-evaluated.Based on the information provided we cannot confirm or further investigate the reported complaint, our investigation remains inconclusive, and a definitive root cause cannot be determined.Specific factors known to contribute to the alleged fault are excessive physical activity levels, unreasonable stress on replacement system, excessive patient weight, trauma to the joint replacement, loosening of components may increase production of wear particles and accelerate damage to the bone.Should the devices or additional information be received, the complaint will be reopened.Based on this investigation, the need for corrective and preventative actions is not indicated.Internal complaint reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKN BIRMINGHAM HIP RESURFACING (BHR) CUP (UNKN TYPE)
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK   CV31 3HL
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15242812
MDR Text Key298073715
Report Number3005975929-2022-00456
Device Sequence Number1
Product Code JDG
Combination Product (y/n)N
Reporter Country CodeIT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/26/2022
Initial Date FDA Received08/17/2022
Supplement Dates Manufacturer Received09/16/2022
Supplement Dates FDA Received09/29/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Other;
Patient Age76 YR
-
-