Siemens has concluded its investigation.A review of the rp500 sodium and potassium sensors' reagent millivolt, sensor slope and raw response curves as well as calibration history for this system indicate that the sodium and potassium sensor were functioning as expected.The observed differences in the sodium and potassium values between the rp500 and the chemistry lab measurement may be affected by differences in methodologies and sample matrixes between the point of care device and the chemistry system.The literature suggests that for sodium, serum samples typically run higher than whole blood samples.The different methodologies (direct vs indirect) will play as a factor since indirect is affected by the amount of protein and lipids dissolved in the sample.A definitive root-cause for the discrepant values could not be determined.
|