Updated fields: b4, d9(device available for eval), g3, g6, h2, h3, h4, h6(investigation type, investigation findings & investigation conclusions), h10 a getinge field service engineer (fse) was dispatched to investigate the issue.In order to fix the issue, the fse performed a 30 psi pressure transducer calibration.After doing so, the calibration passed and the unit was able to start up and run normally.The fse performed all functional and safety checks to factory specifications.The unit was returned to the customer and cleared for clinical use.
|