MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE I PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE I PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-45
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/04/2022
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that during power up, the cardiosave intra-aortic balloon pump (iabp) unit failed electrical test #4.There was no patient involvement.
 
Event Description
N/a.
 
Manufacturer Narrative
Updated fields: b4, d9(device available for eval), g3, g6, h2, h3, h4, h6(investigation type, investigation findings & investigation conclusions), h10 a getinge field service engineer (fse) was dispatched to investigate the issue.In order to fix the issue, the fse performed a 30 psi pressure transducer calibration.After doing so, the calibration passed and the unit was able to start up and run normally.The fse performed all functional and safety checks to factory specifications.The unit was returned to the customer and cleared for clinical use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE I PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key15248878
MDR Text Key305171180
Report Number2249723-2022-02096
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108421
UDI-Public10607567108421
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-45
Device Catalogue Number0998-00-0800-45
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/04/2022
Initial Date FDA Received08/18/2022
Supplement Dates Manufacturer Received06/05/2023
Supplement Dates FDA Received06/06/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/21/2021
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-