Model Number ORT200 |
Device Problem
Unintended Movement (3026)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/22/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The imris customer service engineer was on-site and traced the issue observed with the rotational locking mechanism to the actuator component, which was removed and replaced with a new actuator.The actuator removed from the table was observed to function intermittently and is being returned to the manufacturer for further analysis.The table was functionally tested after component replacement and confirmed to function as intended.
|
|
Event Description
|
During a service visit at a customer site, an imris customer service engineer observed that the intraoperative sliding doors would not open due to the operating room table's rotational position not locking correctly.Upon further assessment the engineer found intermittent function of the actuator component within the table's rotational locking mechanism, and confirmed it had not locked the table's rotational position as intended.There was no patient involvement at the time.
|
|
Manufacturer Narrative
|
The actuator component was returned to the manufacturer and assessed by the component supplier, who confirmed the reported malfunction due to a switch within the actuator.Internal corrective action has been initiated to further address this issue.
|
|
Search Alerts/Recalls
|