• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO BCT SPECTRA OPTIA; SPECTRA OPTIA EXCHANGE SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO BCT SPECTRA OPTIA; SPECTRA OPTIA EXCHANGE SET Back to Search Results
Catalog Number 10226
Device Problems Patient-Device Incompatibility (2682); Adverse Event Without Identified Device or Use Problem (2993); Insufficient Information (3190)
Patient Problem Itching Sensation (1943)
Event Date 07/03/2022
Event Type  Injury  
Event Description
The customer reported that a patient appeared to have an anaphylactic reaction during a tpe procedure.Per the customer, the patient felt slight pruritus on the face at 11:00.The operator gave the patient 10% calcium gluconate 20ml orally and 3mg hexadecadrol per physician order by intravenous injection.The disposable set is not available for return for evaluation.This product is not available within the us, but this report is being filed due to an alleged failure that could occur on a similarly marketed device platform cleared for use by the fda.
 
Manufacturer Narrative
Investigation is in process.A follow up report will be provided.
 
Manufacturer Narrative
Investigation is in process.A follow up report will be provided.
 
Event Description
The customer reported that a patient appeared to have an anaphylactic reaction during a tpe procedure.Per the customer, the patient felt slight pruritus on the face at 11:00.The operator gave the patient 10% calcium gluconate 20ml orally and 3mg hexadecadrol per physician order by intravenous injection.The disposable set is not available for return for evaluation.This product is not available within the us, but this report is being filed due to an alleged failure that could occur on a similarly marketed device platform cleared for use by the fda.
 
Manufacturer Narrative
This report is being filed to provide additional information: investigation: a review of the device history record (dhr) for this unit showed no irregularities during manufacturing that were relevant to this issue.The run data file (rdf) was analyzed for this event.There were no signals or alarms in the rdf that would indicate issues during the run, specifically the amount of ac delivered.The commanded ac pump speed and the measured ac pump speed were accurate and did not show any discrepancy in pump volume.According to therapeutic apheresis: a physician's handbook, adverse events occur during therapeutic procedures with a frequency of 4.8%.Symptoms of allergic reactions may include hives, dyspnea, wheezing, burning eyes, tachycardia, hypotension, and or facial swelling and flushing.Mild reactions can be treated with diphenhydramine administered through an iv.Transient hypocalcemia associated with apheresis is usually well tolerated.Symptoms often show as parasethesia (tingling) but patients may also experience unusual taste, nausea, lightheadedness, shivering, and tremors.Severe hypocalcemia may also cause muscle contractions and can progress to tetany and seizures if hypocalcemia escalates and is not corrected.A disposable complaint history search was performed for this lot and found a total of 3 occurrences reported from the same customer (in this complaint record) for similar issues worldwide.Mdr 1722028-2022-00276 and mdr 1722028-2022-00278.The third occurrence was not reportable.Root cause: a root cause assessment was performed for this complaint.Based on customer's statements about the patient reactions possible causes include but are not limited to: an allergy to replacement solution and/or patient sensitivity to eto patient disease state and/or patient sensitivity to anticoagulant.
 
Event Description
The customer reported that a patient appeared to have an anaphylactic reaction during a tpe procedure.Per the customer, the patient felt slight pruritus on the face at 11:00.The operator gave the patient 10% calcium gluconate 20ml orally and 3mg hexadecadrol per physician order by intravenous injection.The disposable set is not available for return for evaluation.This product is not available within the us, but this report is being filed due to an alleged failure that could occur on a similarly marketed device platform cleared for use by the fda.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRA OPTIA
Type of Device
SPECTRA OPTIA EXCHANGE SET
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
Manufacturer (Section G)
TERUMO BCT
10810 w. collins ave
lakewood CO 80215
Manufacturer Contact
scot hilden
10810 w. collins ave
lakewood, CO 80215
MDR Report Key15252789
MDR Text Key298166795
Report Number1722028-2022-00278
Device Sequence Number1
Product Code LKN
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date 08/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/01/2023
Device Catalogue Number10226
Device Lot Number2110013230
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/26/2022
Initial Date FDA Received08/18/2022
Supplement Dates Manufacturer Received05/04/2023
05/19/2023
Supplement Dates FDA Received05/08/2023
05/31/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/01/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
Patient Age24 YR
Patient SexFemale
Patient Weight65 KG
-
-