A steris service technician arrived onsite to inspect the harmony led 585 surgical lighting system and found that multiple covers to the lighting system were cracked, and the joint cover was missing.This is indicative of facility personnel bumping the lighting system into other pieces of equipment.The harmony led 585 surgical lighting system operator manual states (5-7), "caution - possible equipment damage hazard: do not bump lightheads into walls or other equipment.Always use handles when positioning lighthead during surgical procedures, or when cleaning or servicing the lighting system.Do not bump lightheads into walls or other equipment.Always use handles when positioning lighthead during surgical procedures, or when cleaning or servicing the lighting system." the technician replaced the damaged covers and missing joint cover, tested the lighting system, confirmed it to be operating according to specification, and returned it to service.The technician counseled user facility personnel on the proper use and operation of the harmony led 585 surgical lighting system, specifically the importance of not bumping the lighting system into other pieces of equipment.No additional issues have been reported.
|