• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO BCT TRIMA ACCEL; TRIMA ACCEL AUTOMATED BLOOD COLLECTION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO BCT TRIMA ACCEL; TRIMA ACCEL AUTOMATED BLOOD COLLECTION SYSTEM Back to Search Results
Model Number 3810002
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/12/2022
Event Type  malfunction  
Manufacturer Narrative
The customer confirmed the iv pole safety collar was installed prior to the reported incident.A terumo bct service engineer visited the customer site and found that the button actuator was not making contact with the locking mechanism.The service engineer adjusted the button actuator and returned the device to service.The device serial number history report indicates no further related issues have been reported for this device.All available service history was reviewed for this device with no issues related to the reported condition identified.The device was first installed on (b)(6) 2022.Root cause: a root cause assessment was performed for this complaint.The root cause was determined to be the need for a button actuator adjustment.
 
Event Description
The customer reported that the iv pole does not stay up and drops.There was not a patient or operator involved at the time the iv pole fell, therefore patient information is not reasonably known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIMA ACCEL
Type of Device
TRIMA ACCEL AUTOMATED BLOOD COLLECTION SYSTEM
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
Manufacturer (Section G)
TERUMO BCT
10810 w. collins ave
lakewood CO 80215
Manufacturer Contact
scot hilden
10810 w. collins ave
lakewood, CO 80215
MDR Report Key15253025
MDR Text Key304032893
Report Number1722028-2022-00280
Device Sequence Number1
Product Code GKT
UDI-Device Identifier05020583810006
UDI-Public05020583810006
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
BK190332
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3810002
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/26/2022
Initial Date FDA Received08/18/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-