Roughly 60% of lifepak 20e defibrillators hospital-wide are showing signs of stress-cracking in the form of hairline cracks along the top right side of the front panel case.The problem was originally discovered when two lifepak 20e's came in from the or with almost identical cracking patterns, the discovery of which prompted an investigation into the other likepack 20e's in the hospital, where 44 of 71 saw similar hairline cracks, or the beginnings of such cracking.All defibrillators were manufactured between october/2020 and december/2020.The widespread nature of such failure indicates a possible manufacturing flaw, a conclusion made after confirming the cleaning solutions used on these devices followed that of the cleaning requirements laid out in the lifepak 20e user's manual.Manufacturer response for defibrillator, lifepak (per site reporter).An account representative for the manufacturer was contacted in early august and we are still waiting for a response regarding this finding.
|