FISHER & PAYKEL HEALTHCARE LTD OPTIFLOW JUNIOR INTERFACE; BTT Back to Search Results
Model Number OPT316
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/02/2022
Event Type  malfunction  
Event Description
A distributor in japan on behalf of a healthcare facility reported via a fisher and paykel (f&p) representative that the tubing of an opt316 optiflow junior nasal cannula was found detached from nasal prongs during use.There were no reported patient consequences.
 
Manufacturer Narrative
(b)(4).Fisher & paykel healthcare are currently in the process of retrieving further information regarding this complaint.We will provide a follow up report upon the completion of our investigation.
 
Event Description
A distributor in japan on behalf of a healthcare facility reported via a fisher and paykel (f&p) representative that the tubing of an (b)(6)optiflow junior nasal cannula was found detached from nasal prongs during use.There were no reported patient consequences.
 
Manufacturer Narrative
(b)(4).The optiflow junior nasal cannula is designed specifically for the delicate anatomical features and flow requirements of neonatal and paediatric patients.It features adhesive pads (wigglepads) to maintain cannula stability on the patient's cheeks, soft-touch nasal prongs and breathable kink-proof and crush-resistant flexible tubing.Method: the complaint opt316 optiflow junior nasal cannula was not returned to fisher & paykel healthcare (f&p) for evaluation.Our investigation is thus based on the information and photograph provided by the customer and our knowledge of the product.Result: visual inspection of the provided photograph revealed that the tubing of the complaint device was detached from the right cannula tube joint.Conclusion: without the complaint device, we are unable to determine the cause of the reported event.All optiflow junior cannula are 100% leak and occlusion tested after final assembly and any cannula that fails is discarded.In addition, samples are taken hourly from each run and pull tested to check glue joint strength at the cannula/tube joint, as well as the swivel grip joint.If there are any failures the entire batch is put aside for further investigation.The subject opt316 optiflow junior nasal cannula would have met the required specification at the time of production.The user instructions which accompany the opt316 optiflow junior nasal cannula show in pictorial format the correct placement and fitting of the cannula and also warn: - ensure that all connections are secure during use.Check cannula is undamaged and that the flow path is maintained.Under excessive load, the cannula may disconnect to prevent forces being transferred to the patient.- appropriate patient monitoring must be used at all times.Failure to monitor the patient may result in loss of therapy, serious injury or death.- do not stretch or crush tube.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLOW JUNIOR INTERFACE
Type of Device
BTT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
17400 laguna canyon road,
suite 300
irvine, CA 92618
9494534002
MDR Report Key15280408
MDR Text Key298595696
Report Number9611451-2022-00786
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOPT316
Device Catalogue NumberOPT316
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/04/2022
Initial Date FDA Received08/23/2022
Supplement Dates Manufacturer Received11/03/2022
Supplement Dates FDA Received12/01/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-