FISHER & PAYKEL HEALTHCARE LTD OPTIFLOW + ADULT NASAL CANNULA; BTT Back to Search Results
Model Number OPT944
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/18/2022
Event Type  malfunction  
Event Description
A distributor reported on behalf of a healthcare facility in japan that the tubing of two opt944 optiflow + adult nasal cannulas had detached from the 3-way connector during patient use.There were no reported patient consequences.
 
Manufacturer Narrative
(b)(4).Fisher & paykel healthcare (f&p) is currently in the process of completing our investigation.We will provide a follow up report upon completion of investigation.Product background: the opt944 optiflow + adult nasal cannula is a nasal cannula patient interface for delivery of humidified respiratory gases, including those who are receiving nasal high flow therapy (nhf).Optiflow + interfaces are designed for use with the airvo 2 series of humidification devices.Nhf therapy and the airvo 2 device are intended for use with spontaneously breathing patients who would benefit from receiving high flow warmed and humidified respiratory gases.The airvo 2 device should not be used for life support purposes, and appropriate patient monitoring must be used at all times.The cannula consists of a lightweight delivery tube connected to a rigid plastic base and soft nasal prongs (nasal interface).This allows the device to be light weight and durable.The interface is held in place by a head strap and features a head strap clip which works in tandem with the tubing clip (attaches to the patient's clothing/bedding) to support the weight of the circuit and prevent the cannula being dislodged.
 
Event Description
A distributor reported on behalf of a healthcare facility in japan that the tubing of two opt944 optiflow + adult nasal cannulas had detached from the 3-way connector during patient use.There were no reported patient consequences.
 
Manufacturer Narrative
(b)(4).Product background: the opt944 optiflow + adult nasal cannula is a nasal cannula patient interface for delivery of humidified respiratory gases, including those who are receiving nasal high flow therapy (nhf).Optiflow + interfaces are designed for use with the airvo 2 series of humidification devices.Nhf therapy and the airvo 2 device are intended for use with spontaneously breathing patients who would benefit from receiving high flow warmed and humidified respiratory gases.The airvo 2 device should not be used for life support purposes, and appropriate patient monitoring must be used at all times.The cannula consists of a lightweight delivery tube connected to a rigid plastic base and soft nasal prongs (nasal interface).This allows the device to be light weight and durable.The interface is held in place by a head strap and features a head strap clip which works in tandem with the tubing clip (attaches to the patient's clothing/bedding) to support the weight of the circuit and prevent the cannula being dislodged.Method: one of the two complaint opt944 optiflow + adult nasal cannulas was received at fisher & paykel healthcare (f&p) in new zealand for investigation, where it was visually inspected.Our investigation is based on our evaluation of the returned device, the information provided by the customer, previous investigations of similar complaints, and our knowledge of the product.Results: visual inspection revealed the tubing of the returned opt944 optiflow + adult nasal cannula was detached from the 3-way connector and the right hook was detached from the interface.It was further reported by the healthcare facility that the head strap clip and tubing clip were not used to secure the cannula.Conclusion: our investigation was unable to determine the cause of the observed damage to the opt944 optiflow + adult nasal cannula.However, based on our knowledge of the product and the type of damage observed, the reported damage is likely to have been caused by the tubing being subjected to excessive force during use.It should also be noted that the head strap clip and tubing clip were not used to secure the cannula.Manufacturing controls include inspections during production for visual defects to the optiflow + tubing and the swivel, including cracks, tears, moulding defects, contamination, inclusions, discoloration and stretching or deformation.The optiflow + tubing is also inspected for any assembly defects, including confirmation the swivel and 3-way connector are connected with the correct engagement.Any product that fails the visual inspection is disposed of.The subject opt944 optiflow + adult nasal cannula would have met the required specifications.The user instructions which accompany the opt944 optiflow + adult nasal cannula show in pictorial format the correct placement and fitting of the cannula, including ensuring the headstrap clip and the tubing clip are appropriately attached.The user instructions also warn: "ensure head strap clip is attached, to prevent cannula from being pulled out of the nares."."cannula can become unattached if not used with the head strap clip."."attach tubing clip to clothing/bedding to prevent cannula from pulling off face."."appropriate patient monitoring must be used at all times.Failure to monitor the patient may result in loss of therapy, serious injury or death."."do not crush or stretch tube, to prevent loss of therapy."."failure to use the set-up described above can compromise performance and affect patient safety.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLOW + ADULT NASAL CANNULA
Type of Device
BTT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
17400 laguna canyon road
suite 300
irvine, CA 92618
9494534002
MDR Report Key15280461
MDR Text Key298492261
Report Number9611451-2022-00789
Device Sequence Number1
Product Code BTT
UDI-Device Identifier09420012429117
UDI-Public(01)09420012429117(10)2101841753(11)211012
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K162553
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOPT944
Device Catalogue NumberOPT944
Device Lot Number2101841753
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/17/2022
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/25/2022
Initial Date FDA Received08/23/2022
Supplement Dates Manufacturer Received10/10/2022
Supplement Dates FDA Received10/11/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/12/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
F&P 900PT561 HEATED BREATHING TUBE AND CHAMBER KIT; F&P 900PT561 HEATED BREATHING TUBE AND CHAMBER KIT; F&P PT101 AIRVO 2 HUMIDIFIER; F&P PT101 AIRVO 2 HUMIDIFIER
Patient SexMale
-
-