• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AOMORI OLYMPUS CO., LTD. SINGLE USE INJECTOR; INJECTOR AND SHEATHSET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AOMORI OLYMPUS CO., LTD. SINGLE USE INJECTOR; INJECTOR AND SHEATHSET Back to Search Results
Model Number NM-401L-0525
Device Problem Difficult to Flush (1251)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The customer (physician) reports during an injection of steroid into the vocal cords procedure using a single use injector, two injectors would not inject fluid (there was resistance).The procedure was completed with a similar device.There were no adverse effects to the patient.Case with patient identifier (b)(6) reports injector one of two.Case with patient identifier (b)(6) reports injector two of two.
 
Manufacturer Narrative
The device referenced in this report was not returned to olympus for evaluation.The definitive cause of the user's experience cannot be determined at this time.The investigation is ongoing.This report will be updated upon completion of the investigation or upon receipt of additional relevant information.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, since the device was not returned the exact cause could not be determined.However, a similar complaint reporting ¿unable to inject liquid into the target tissue,¿ occurred due to the compressive bucking on the needle tube.It is unlikely that a device with a defective needle was shipped, as the device undergoes full inspection for appearance, needle operation and injection.Therefore, the compressive buckling on the needle tube was likely caused when the needle was extended due to great friction between the outer tube and the needle.The event can be detected/prevented by following the instructions for use (ifu) which state: ¿the operator of this instrument must be physician or medical personnel under the supervision of a physician and must have received sufficient training in clinical endoscopic technique.This manual does not explain or discuss clinical endoscopic procedures.¿ ¿straighten out the instrument before inspecting it.The instrument can be damaged if it is coiled while the handle is operated.¿ ¿operate the slider slowly, otherwise the tube could buckle.¿ ¿when inserting the instrument into the endoscope, retract the needle into the sheath, hold the instrument close to the biopsy valve, and keep it as straight as possible relative to the biopsy valve.Otherwise, the instrument could be damaged.¿ ¿insert the instrument slowly.Abrupt insertion could damage the endoscope and/or instrument.¿ ¿stop using the instrument if the insertion portion bends excessively during use.This could result in malfunction, such as failing to extend the needle or inject a fluid.¿ olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE INJECTOR
Type of Device
INJECTOR AND SHEATHSET
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer Contact
masaharu hirose
2-248-1 okkonoki
kuroishi-shi, aomori 036-0-357
JA   036-0357
426422891
MDR Report Key15294774
MDR Text Key305353714
Report Number9614641-2022-00209
Device Sequence Number1
Product Code FBK
UDI-Device Identifier04953170260094
UDI-Public04953170260094
Combination Product (y/n)N
Reporter Country CodeLE
PMA/PMN Number
K902736
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberNM-401L-0525
Device Lot Number14V
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/28/2022
Initial Date FDA Received08/25/2022
Supplement Dates Manufacturer Received09/12/2022
Supplement Dates FDA Received10/04/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-