• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY LLC AUTOMATE TOOL REFILL; INSTRUMENTS, DENTAL HAND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY LLC AUTOMATE TOOL REFILL; INSTRUMENTS, DENTAL HAND Back to Search Results
Catalog Number 663100
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
While no serious injury resulted in this event, if this malfunction recurred, it could cause or contribute to a serious injury or require medical or surgical intervention to preclude such.This event.Therefore; is reportable per 21cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
 
Event Description
In this event it is reported that automate tool broke during use while in patient's mouth.No injury occurred.
 
Manufacturer Narrative
Returned product was 1 automate iii handle and 1 flexshaft date coded 0721 (manufactured july 2021) with weld failure causing the tip or burr component to be removed from the flexshaft assembly thus causing the product to not function properly.A capa has been opened to address weld failures for product manufactured by sarasota since december 2020 (date code 1220).Dhr review/retain evaluation cannot be conducted without batch information.Complaint is considered substantiated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOMATE TOOL REFILL
Type of Device
INSTRUMENTS, DENTAL HAND
Manufacturer (Section D)
DENTSPLY LLC
38 west clarke avenue
milford DE 19963
Manufacturer (Section G)
DENTSPLY LLC
38 west clarke avenue
milford DE 19963
Manufacturer Contact
hannah seevaratnam
221 west philadelphia st.
york, PA 17401
7178457511
MDR Report Key15303872
MDR Text Key303275193
Report Number2515379-2022-00053
Device Sequence Number1
Product Code DZN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 09/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number663100
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date08/24/2022
Initial Date Manufacturer Received 08/24/2022
Initial Date FDA Received08/26/2022
Supplement Dates Manufacturer Received08/24/2022
Supplement Dates FDA Received09/16/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Type of Device Usage A
Patient Sequence Number1
-
-