• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL INC. 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL INC. 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100 B
Device Problem Temperature Problem (3022)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/02/2022
Event Type  malfunction  
Event Description
It was reported to vyaire medical device overheated and stopped oscillating during use on a patient on the 3100 b ventilator.At this time, there is no information regarding patient harm associated with the reported event.
 
Manufacturer Narrative
At this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Manufacturer Narrative
"results of investigation: vyaire field service representative went onsite to evaluate the device.Technician troubleshot and tested unit.Left oscillator running for one hour with settings that were on the unit at arrival with no problem.Tested oscillator for another hour with the performance check settings and unit worked within manufacturer specifications.Unit is ready for use.".Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL INC.
26125 n. riverwoods blvd.
mettawa IL 60045
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
erika bonilla
510 technology dr
irvine, IL 92618
7149227837
MDR Report Key15307478
MDR Text Key305414595
Report Number2021710-2022-16542
Device Sequence Number1
Product Code LSZ
UDI-Device Identifier10846446003543
UDI-Public(01)10846446003543(11)20100205
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100 B
Device Catalogue Number770155
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/02/2022
Initial Date FDA Received08/26/2022
Supplement Dates Manufacturer Received07/28/2023
Supplement Dates FDA Received12/15/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/02/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-