Model Number P0819A
Device Problem Electrical Power Problem (2925)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/16/2022
Event Type  malfunction  
Manufacturer Narrative
The hillrom technician found the unit end power cord on envella® bed was burnt.Per the hillrom service manual, perform annual preventive maintenance procedures to make sure all envella® bed components are functioning as originally designed.The user manual instructs that power cords require inspection routinely to ensure no damage has occurred during use.The manuals also include warnings such as incorrect use or handling may result in damage to the power cord.They also instruct that if damage has occurred remove immediately, and to properly remove the power cord from the electrical outlet during transfer or movement of the device.A search of the hillrom maintenance records showed hillrom performed preventative maintenance on this bed sep 15, 2021.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced the power cord to resolve the issue.Based on this information, no further action is required.
 
Event Description
Hillrom received a report from a hillrom technician stating the end of the power cord on envella bed was burnt.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENVELLA BED
Type of Device
BED, AIR FLUIDIZED
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
melrita gilliam
1069 state route 46 east
batesville, IN 47006
3127289851
MDR Report Key15310544
MDR Text Key298778686
Report Number1824206-2022-00382
Device Sequence Number1
Product Code INX
UDI-Device Identifier00887761013834
UDI-Public010088776101383411190117
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 08/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberP0819A
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/16/2022
Initial Date FDA Received08/29/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/07/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-