SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX TRACHEOSTOMY TUBE; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Device Problem Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/01/2020
Event Type  malfunction  
Event Description
It was reported that the device was disconnecting while the customer was using a filter cover.No patient injury reported.
 
Manufacturer Narrative
No product was returned therefore the customer complaint was not confirmed.However, we connected the device connector and flat filter in our stock, and wrapped it in the filter cover and connected the disposable infusion pump made by another company.As a result of confirming the connection points, it was confirmed that the gap between the connectors was smaller than that of the luer type connector.However, the reported event was not confirmed as the filter cover did not open when connecting the infusion pump made by another company.No lot number was provided; therefore, device history record review could not be completed.No product information has been provided to date.This remediation mdr was generated under protocol b10009406, as a result of warning letter cms# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX TRACHEOSTOMY TUBE
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
aichi pref.
minneapolis, MN 55442
MDR Report Key15310823
MDR Text Key305356901
Report Number3012307300-2022-16492
Device Sequence Number1
Product Code BTO
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 08/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/13/2022
Initial Date FDA Received08/29/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-