RESPIRONICS CALIFORNIA, LLC RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problems
Mechanical Problem (1384); Blocked Connection (2888)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/09/2022 |
Event Type
malfunction
|
Event Description
|
The customer reported that the device alarms patient circuit occluded, and the blower was not spinning.There was no patient involvement when the issue occurred.No patient or user harm reported.Per the diagnostics performed by the engineer, the remote service engineer (rse) advised the customer that the device may have a faulty blower assembly or a faulty motor controller pcb.Further information has been requested.If additional information becomes available at a later date, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
Multiple attempts to retrieve device repair or diagnostic information has yielded no response from the customer.It is unknown if any parts or repair has been conducted.
|
|
Search Alerts/Recalls
|
|
|