Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Should additional information prompt us to alter or supplement any information or conclusions contained in this report, a follow-up report will be submitted.
|
According to the available information, the trocar for the patient's left side was loaded with the dynamic anchor after the static anchor had already been placed.The surgeon placed the trocar in the incision but pulled it back out to check the placement of the anchor on the trocar.The trocar was bent, so the surgeon tried to straighten the tip with his fingers, but ultimately broke the trocar tip.The surgeon commented that the trocar tip appeared weakened.The static anchor was removed and a second altis sling system was opened and successfully placed.
|