Lot Number 200121
Device Problem Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/01/2020
Event Type  malfunction  
Manufacturer Narrative
A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.The device was returned for evaluation.Visual and functional tests were performed.It was confirmed that when connecting the flat filter and the epifuse connector, the collar easily came off and moved to the epifuse connector.It was observed that the retention edge of the collar was slightly crushed.The root cause of the reported issue was not possible to identify the cause of the event.The supplier and the manufacturing department will be notified, record this event in the database, and monitor the future occurrence situation.Device information is unknown.No product information has been provided to date.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Event Description
It was reported that when turning the collar of the flat a little too far, it slipped off from the connector.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTINUOUS EPIDURAL ANESTHESIA TRAY
Type of Device
ANESTHESIA CONDUCTION KIT
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
10 bowman dr.
keene NH 03431
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
10 bowman dr.
keene NH 03431
Manufacturer Contact
jim vegel
6000 nathan lane north
3-10-15, hakata station east,
minneapolis, MN 55442
MDR Report Key15322389
MDR Text Key305180612
Report Number3012307300-2022-16832
Device Sequence Number1
Product Code CAZ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K781052
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 08/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Lot Number200121
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/10/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/02/2020
Initial Date FDA Received08/30/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-