• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH EARLYVUE VS30 VITALS MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH EARLYVUE VS30 VITALS MONITOR Back to Search Results
Model Number 863380
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/03/2022
Event Type  malfunction  
Manufacturer Narrative
A follow-up report will be submitted upon completion of the investigation.
 
Event Description
The customer reported that the device was tipped over and screen is shattered.The device was not in use.
 
Event Description
The customer reported that the device was tipped over and screen is shattered.The device was not in use on a patient at the time of event, there was no patient involvement.The device was returned to philips bench repair for evaluation.The bench repair technician observed the shattered screen.Multiple good faith efforts were made to obtain additional information on how the device was tipped over, but there were no response.Therefore, we are considering this to be a malfunction of cause unknown.The lcd assembly was replaced and the device was returned to the customer's site.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EARLYVUE VS30 VITALS MONITOR
Type of Device
EARLYVUE VS30 VITALS MONITOR
Manufacturer (Section D)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
jeanne ahearn
222 jacobs st
cambridge, MA 02141
6172455900
MDR Report Key15322700
MDR Text Key301027230
Report Number9610816-2022-00469
Device Sequence Number1
Product Code DSJ
UDI-Device Identifier00884838091412
UDI-Public00884838091412
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K190624
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number863380
Device Catalogue Number863380
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/03/2022
Initial Date FDA Received08/30/2022
Supplement Dates Manufacturer Received10/19/2022
Supplement Dates FDA Received11/02/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured11/23/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-