• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION STONE CONE; DISLODGER, STONE, FLEXIBLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION STONE CONE; DISLODGER, STONE, FLEXIBLE Back to Search Results
Model Number M0063903200
Device Problem Peeled/Delaminated (1454)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/24/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4) the returned stone cone device was analyzed, and a visual evaluation observed that the distal end of the blue/green heat sheath was separated from the distal stop.Based on all available information, it is most likely that the cone was being tested through the distal end causing the green/blue sheath to separate from the distal stop making the cone easy to open.The instructions for use (ifu) states that prior to use, ensure that the coil is working properly by advancing the sheath over the coil to the positive stop and then retracting the sheath to open the coil.The sheath of the device should be straight during testing.Therefore, the most probable root cause is unintended use error.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
 
Event Description
It was reported to boston scientific corporation that a stone cone nitinol retrieval coil was used in the ureter during a lithotomy under ureteroscopy procedure performed on (b)(6) 2022.During preparation, it was found that the basket would open too easily with a pull.The procedure was completed with another stone cone.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.The investigation results revealed the coil/cone peeled/sheared; therefore, this is now an mdr reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STONE CONE
Type of Device
DISLODGER, STONE, FLEXIBLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
LAKE REGION MEDICAL
31c butterfield trail
el paso TX 79906
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15322867
MDR Text Key305425927
Report Number3005099803-2022-04902
Device Sequence Number1
Product Code FGO
UDI-Device Identifier08714729430223
UDI-Public08714729430223
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 08/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/14/2023
Device Model NumberM0063903200
Device Catalogue Number390-320
Device Lot Number0005225784
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/13/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/11/2022
Initial Date FDA Received08/30/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/14/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age56 YR
Patient SexMale
Patient Weight64 KG
-
-