Model Number A1059
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A facility reported that the ratchet mechanism on the mayfield modified skull clamp (b)(4) jumps when opening and closing and the pressure pin is losing pressure.The facility has not reported if there was patient involvement, injury, or surgical delay relative to this event.
 
Manufacturer Narrative
Mayfield modified skull clamp (a1059) was returned for evaluation: failure analysis: the complaint is confirmed via inspection of the unit.The investigation showed that the lock had rotational and lateral movement and a residue buildup was present.The ratchet extension was not moving smoothly in the body casting.General maintenance and cleaning performed; new components were added to replace worn internal parts.Root cause analysis: probable root cause includes rough handling and routine wear of the device.Unit required maintenance and replacement of worn components.No further investigation is required based on the acceptability of risk and no adverse trends were identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD MODIFIED SKULL CLAMP
Type of Device
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key15324423
MDR Text Key300446353
Report Number3004608878-2022-00180
Device Sequence Number1
Product Code HBL
UDI-Device Identifier10381780253457
UDI-Public10381780253457
Combination Product (y/n)N
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA1059
Device Catalogue NumberA1059
Device Lot Number194
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/11/2022
Initial Date FDA Received08/30/2022
Supplement Dates Manufacturer Received10/12/2022
Supplement Dates FDA Received10/17/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-