A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product sample was received for evaluation with a photo provided.Visual and functional testing were performed.Visual inspection found no visual damage.Functional testing found that splotches on the coating of the device and were most likely occurred during the coating process.The root cause of the reported issue was found to be th splotches on the coating.D5 operator of device is patient/consumer.No information is provided to date.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# 617147.
|