MEDTRONIC, INC. CARELINK ENCORE PROGRAMMER; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 29901A
Device Problems Computer Software Problem (1112); Device Difficult to Program or Calibrate (1496); Operating System Becomes Nonfunctional (2996)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/25/2022
Event Type  malfunction  
Manufacturer Narrative
Analysis was unable to confirm the customer comment that the programmer crashed multiple times to a black screen.It was noted that the programmer stylus was not in alignment with the cursor.It was also noted that part of the display was separating and was moving when touched.The computer platform was replaced.The programmer hard drive was reconfigured and the programmer software was reloaded and updated.All found defective parts were replaced and all other identified issues were resolved.The instrument then passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the programmer crashed multiple times to a black screen with a diagnostic message outside of patient session.The programmer was returned to service and subsequently tested out of specification during manufacturer's analysis.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK ENCORE PROGRAMMER
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key15327462
MDR Text Key305371887
Report Number2182208-2022-02814
Device Sequence Number1
Product Code KRG
UDI-Device Identifier00643169763548
UDI-Public00643169763548
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 08/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number29901A
Device Catalogue Number29901A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/18/2022
Initial Date Manufacturer Received 08/25/2022
Initial Date FDA Received08/31/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/12/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-