MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL) Back to Search Results
Catalog Number 8065977763
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problems Conjunctivitis (1784); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/08/2022
Event Type  malfunction  
Event Description
A nurse reported that following implantation of an intraocular lens (iol) a foreign material found on the back of implant, surgeon unsure if it was from the cartridge or something else.Lens was loaded and foreign material appeared on back of the lens, surgeon removed this with irrigation/aspiration.Additional information was requested and received stating the foreign material that is actually still lodged in the patients eye and did not affect the patients vision.Additional information was received stating that the patient is doing well, nothing to report, healing as per normal.
 
Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
Photo was provided.There is no context visible.Magnified photo of a particle.Elongated material, may appear tinted due to lighting.Appears to be free floating.No determination as to the nature of origin of the material can be made from the photo.Video: a video clip was provided which shows the pictured material in the pupil area.No determination as to the nature of origin of the material can be made from the video.The manufacturer internal reference number is: (b)(4).
 
Event Description
Additional information was requested and received vision is good but she still has a greenish foreign body behind the lens.The conjunctiva was a bit inflamed as she stopped using her prednefrine prematurely.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D
Type of Device
FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15327754
MDR Text Key298985129
Report Number1119421-2022-01883
Device Sequence Number1
Product Code MSS
UDI-Device Identifier00380659777639
UDI-Public00380659777639
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K063155
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 09/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065977763
Device Lot Number15374829
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/10/2022
Initial Date FDA Received08/31/2022
Supplement Dates Manufacturer Received09/01/2022
Supplement Dates FDA Received09/24/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/08/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-