Model Number N/A
Device Problem Packaging Problem (3007)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/10/2022
Event Type  malfunction  
Event Description
It was reported that the sterile package was damaged.No adverse events have been reported as a result of the malfunction and additional information on the reported event is unavailable.
 
Manufacturer Narrative
(b)(4).Multiple reports have been submitted along with this report 0001825034-2022-01983.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: evaluation of the returned products confirmed damage to the sterile packaging pouch, and there is debris inside the sterile packaging which is consistent with the appearance of porous coating, and foam debris from the foam packaging inside the sterile barrier.Sterility has not been compromised.Device history record was reviewed and no discrepancies relevant to the reported event were found.The likely condition of the products when they left zimmer biomet was conforming to specifications.The root cause of the reported event is likely due to transit damage.This is no longer considered a reportable event.A corrective action has been opened to further investigate this event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC 133 MP TYPE1 PPS SO 14.0
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15329943
MDR Text Key302247744
Report Number0001825034-2022-01982
Device Sequence Number1
Product Code JDI
UDI-Device Identifier00880304514355
UDI-Public(01)00880304514355(17)300623(10)6806716
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K120030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 10/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number51-106140
Device Lot Number6806716
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2022
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/10/2022
Initial Date FDA Received08/31/2022
Supplement Dates Manufacturer Received09/26/2022
Supplement Dates FDA Received10/20/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/23/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-