Siemens filed the initial mdr 1219913-2022-00259 on september 01, 2022.December 05, 2022 additional information: an outside of the united states (ous) customer contacted the siemens customer care center to report high atellica im vitamin b12 (vb12) results for samples from two patients that were considered discordant when compared to the lower repeat results on the other atellica im analyzers at the customer site.Siemens has investigated.Upon review of the data, it was noted that the repeats of the patient samples occurred more than 48 hours later, and testing was performed on alternate vb12 reagent lot 282.Per atellica im vb12 information for use (ifu), 10995437_en rev.02, 2019-08, patient samples must be frozen at less than or equal to -20°c if not tested within 48 hours.The customer uses an aptio high volume sample refrigeration system for sample storage and therefore samples were not frozen and not eligible for vb12 repeat testing past 48 hours.Bio-rad quality control (qc) lot 85290 (level 1 and level 2) was sporadically out of range low according to customer range but within bio-rad package insert ranges during time of patient testing and days following.The quality of the specimen could potentially impact the performance of the atellica im vb12 assay.Cellular debris could potentially be responsible for the higher-than-expected vb12 results.When samples are centrifuged prior to testing and remain upright or allowed to settle for an extended period of time, the vb12 assay can be impacted resulting in lower values.Siemens recommends following all collection tube handling instructions provided by the manufacturer.Siemens performed additional testing for vb12 lot 280 and bio-rad qc lot 85290, which confirmed that this lot of bio-rad qc is performing as expected and within published package insert ranges and close to package insert means.Qc and reagent performance during the study showed no significant degradation over a 4-day period on 3 different packs of vb12 reagent lot 280.No product performance issue was found.Siemens cannot rule out issues with customer bio-rad qc supply, reagent supply or preanalytical handling of qc or patient samples as the cause of qc out of range and the discordant patient results at customer site.Based on the investigation, no product problem was identified.In section h6, an additional type of investigation code was added and the investigation finding, and investigation conclusion codes were updated.
|