Model Number V60
Device Problem Inaccurate Flow Rate (1249)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/15/2022
Event Type  malfunction  
Event Description
The customer reported a ventilator's flow reading was off.The device was not in clinical use at the time the issue was discovered.There was no patient or user harm reported.The device was evaluated by the customer and a philips remote service engineer (rse).The customer reported that they are unable to reproduce the issue.The rse advised the customer to perform a complete performance verification test (pvt) per the service guide with passing results before placing the unit back into service.The customer is to perform a complete pvt with passing results before returning unit to service.Further information has been requested.If additional information becomes available at a later date, a supplemental report will be submitted.
 
Manufacturer Narrative
H10: multiple good faith efforts were made to retrieve device evaluation, repair, and operational status on 08/21/2022, 10/03/2022 and 10/18/2022, however, yielded no response from the customer.It is unknown if any parts or repairs have been conducted.The complaint will be processed for closure.If additional information is received at a later date, the complaint will be reopened, and a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, LLC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
RESPIRONICS CALIFORNIA, LLC
2271 cosmos court
carlsbad CA 92011
Manufacturer Contact
kimberly shelly
2271 cosmos court
carlsbad, CA 92011
7609187300
MDR Report Key15337014
MDR Text Key304817508
Report Number2031642-2022-02276
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/15/2022
Initial Date FDA Received09/01/2022
Supplement Dates Manufacturer Received10/18/2022
Supplement Dates FDA Received10/19/2022
Date Device Manufactured01/10/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-