Model Number DB-1416 |
Device Problems
Premature Discharge of Battery (1057); Inadequate or Insufficient Training (1643)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/04/2022 |
Event Type
malfunction
|
Event Description
|
It was reported the patient experienced premature battery depletion.A report from the field indicated the patient's implantable pulse generator (ipg) was nearing end of life after thirteen months of use.Analysis of the ipg settings and stimulation usage by engineers revealed an energy use index (eui) index of 10.2 with impedances within normal range; indicating an anticipated approximate 36 months of service.Further review of the battery voltages for the ipg revealed an unexpected drop at the time of implant and a more than usual decline in voltage over the thirteen months of service.No further information has been provided despite good faith efforts.
|
|
Event Description
|
It was reported the patient experienced premature battery depletion.A report from the field indicated the patient's implantable pulse generator (ipg) was nearing end of life after thirteen months of use.Analysis of the ipg settings and stimulation usage by engineers revealed an energy use index (eui) index of 10.2 with impedances within normal range; indicating an anticipated approximate 36 months of service.Further review of the battery voltages for the ipg revealed an unexpected drop at the time of implant and a more than usual decline in voltage over the thirteen months of service.No further information has been provided despite good faith efforts.Additional information was received that the patient underwent an explant procedure wherein the ipg was explanted.The patient was doing well postoperatively.
|
|
Event Description
|
It was reported the patient experienced premature battery depletion.A report from the field indicated the patient's implantable pulse generator (ipg) was nearing end of life after thirteen months of use.Analysis of the ipg settings and stimulation usage by engineers revealed an energy use index (eui) index of 10.2 with impedances within normal range; indicating an anticipated approximate 36 months of service.Further review of the battery voltages for the ipg revealed an unexpected drop at the time of implant and a more than usual decline in voltage over the thirteen months of service.No further information has been provided despite good faith efforts.Additional information was received that the patient underwent an explant procedure wherein the ipg was explanted.The patient was doing well postoperatively.
|
|
Manufacturer Narrative
|
Device analysis performed on the returned ipg revealed the primary cell battery reached the end of service (eos) mode prematurely after 13 months of use.Engineers determined that there was a sudden drop in battery voltage at the time of implant which resulted in early ipg battery depletion and shortened the device longevity.Observations from the field indicated that the ipg had not been exposed to magnetic resonance imaging (mri) or any other high energy exposures during implant or shortly after.The cause of the voltage drop could not be determined as no anomalies were found through laboratory analysis.Based on all the available information, the cause of the reported event could not be established.
|
|
Search Alerts/Recalls
|