• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAUSCH + LOMB EASY-LOAD LENS DELIVERY SYSTEM; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAUSCH + LOMB EASY-LOAD LENS DELIVERY SYSTEM; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL) Back to Search Results
Model Number EZ-28V
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Vitreous Loss (2142); Capsular Bag Tear (2639)
Event Date 08/01/2022
Event Type  Injury  
Manufacturer Narrative
Per the reporter, the device has been discarded and is not available for evaluation.The investigation of this event is in progress.A follow-up report will be submitted upon completion of the investigation.
 
Event Description
Reportedly, trailing haptic was noticed bent, after loading the iol into the right eye.The capsular bag was reported broken and there was vitreous loss.The lens was then removed and a vitrectomy was performed.The patient was left aphakic due to the cornea being too cloudy.According to the surgeon, patient outcome will be good as long as corneal edema clears and a backup iol is implanted at a later date.In the physician¿s opinion, the most likely cause of this event is a loading error.
 
Manufacturer Narrative
The device was not returned for evaluation as it was discarded.Although requested, a lot number was not provided; therefore, a device history record (dhr) review could not be performed.The trend analysis, risk analysis and directions for use review were considered acceptable, with the product performing within anticipated rates.Based on the available information, the root cause of this event could not be conclusively determined; however, user-related factors (such as loading or handling techniques) and/or procedural factors (such as lens and inserter interaction) might have caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASY-LOAD LENS DELIVERY SYSTEM
Type of Device
FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)
Manufacturer (Section D)
BAUSCH + LOMB
1400 goodman street
rochester NY 14609
Manufacturer (Section G)
BAUSCH + LOMB
21 north park place blvd.
clearwater FL 33759
Manufacturer Contact
shayan habibi
21 north park place blvd.
clearwater, FL 33759
7277246600
MDR Report Key15349267
MDR Text Key299178345
Report Number0001313525-2022-00112
Device Sequence Number1
Product Code MSS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132481
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberEZ-28V
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/02/2022
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received10/03/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SOFPORT LI61AOR IOL.
Patient Outcome(s) Required Intervention;
Patient Age84 YR
Patient SexMale
-
-