MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION MINICAP; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION MINICAP; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE Back to Search Results
Catalog Number 5C4466P
Device Problem Fitting Problem (2183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/09/2022
Event Type  malfunction  
Manufacturer Narrative
The device was received and is currently awaiting evaluation.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a minicap had a connection issue; further described as ¿when trying to put a minicap on, the minicap would not go on the transfer set¿.This was observed during use of the device for peritoneal dialysis (pd) therapy.It was further reported that ¿the thread on the inside looks totally different¿.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
H4: the lot was manufactured april 08 2022 to april 14 2022.H10:the actual device was received for evaluation.A visual inspection was performed and damage was noted; approximately half of the thread was not properly formed.The wall of the sample showed signs of deformation/damage along the circumference.Functional testing was performed with an in-lab transfer set and the cap could not be secured to the set.The reported condition was verified.The cause of the condition was related to the molding process during manufacturing.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINICAP
Type of Device
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - CLEVELAND
911 highway 61 north
po box 1058
cleveland MS 38732
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key15351566
MDR Text Key302857919
Report Number1416980-2022-04619
Device Sequence Number1
Product Code KDJ
UDI-Device Identifier00085412007694
UDI-Public(01)00085412007694
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K152129
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 10/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2023
Device Catalogue Number5C4466P
Device Lot NumberGD911368
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/19/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/09/2022
Initial Date FDA Received09/02/2022
Supplement Dates Manufacturer Received09/21/2022
Supplement Dates FDA Received10/04/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/08/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
TRANSFER SET
-
-