Siemens has completed the investigaiton.Instrument logs provided only contained the instrument status messages.Without the patient results or decodes available a definitive root cause cannot be determined.Samples that contain intact/not lysed leukocytes have the potential to produce a negative result on the leukocyte esterase pad.It is likely that the leukocytes in the sample that were visible on microscopy were not sufficiently lysed to generate a positive result on the leukocyte esterase pad.Microscopy/sediment analysis detects visible and intact leukocytes, which is different from chemical urinalysis which detects leukocyte esterase.Leukocyte esterase is only present in samples where the leukocytes have lysed and released the analyte into the surrounding specimen.While leukocyte esterase is correlated with the presence of leukocytes in sediment analysis, they are not perfectly predictable of one another.Based off available information the novus appears to be function as intended.
|