• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED CELLEX PHOTOPHERESIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED CELLEX PHOTOPHERESIS SYSTEM Back to Search Results
Catalog Number CLXECP
Device Problems Disconnection (1171); Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/08/2022
Event Type  malfunction  
Manufacturer Narrative
The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction pressure dome leak.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.A batch record review for kit lot l202 was conducted.There were no non-conformances related to the complaint.This lot met all release requirements.A review of kit lot l202 shows no trends.Trends were reviewed for complaint category, pressure dome leak.No trends were detected for this complaint category.The assessment is based on the information available at the time of investigation.No product was returned for investigation; therefore, it could not be determined if the product met specification based solely on the information provided by the customer.Complaints are monitored through tracking and trending.If a trend is detected, further investigation will be conducted.This investigation is now complete.(b)(4).P.T.(b)(6) 2022.
 
Event Description
The customer contacted mallinckrodt to report they experienced a pressure dome leak with their cellex photopheresis kit ("kit") during an extracorporeal photopheresis (ecp) treatment.The customer reported they accidentally pressed stop instead of pause when they were switching from a packed red blood cell (prbc) bag to the patient.The customer stated when resuming the procedure, the return pressure dome disconnected from the pressure sensor.The customer stated the patient return line was clamped at the time the pressure dome leak occurred.No blood had been drawn from the patient.The ecp treatment was aborted and residual blood within the kit was not returned to the patient.The customer reported the patient was in stable condition and would start a new ecp treatment using a new prbc bag.The customer did not return the product for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELLEX PHOTOPHERESIS SYSTEM
Type of Device
CELLEX PHOTOPHERESIS SYSTEM
Manufacturer (Section D)
MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED
blanchardstown, dublin D15 T X2V
EI  D15 TX2V
Manufacturer (Section G)
MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED
college business & tech park
cruiserath road
blanchardstown, dublin D15 T X2V
EI   D15 TX2V
Manufacturer Contact
megan vernak
shelbourne building
53 frontage rd suite 300
hampton, NJ 08827
MDR Report Key15363568
MDR Text Key305341365
Report Number3013428851-2022-00066
Device Sequence Number1
Product Code LNR
Combination Product (y/n)Y
Reporter Country CodeUK
PMA/PMN Number
P860003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 09/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/01/2024
Device Catalogue NumberCLXECP
Device Lot NumberL202
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/08/2022
Initial Date FDA Received09/06/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/07/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-