C.R. BARD, INC. (COVINGTON) -1018233 LATEX FOLEY CATHETER; UNKNOWN FOLEY CATHETER Back to Search Results
Device Problem Dent in Material (2526)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/31/2022
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that the customer had a leakage issue with the foley catheter.Stated that the catheter inflated to one side and was causing this issue.And their doctor thought it was bladder spasm and tried over and under inflating the catheter but did not solve this.
 
Manufacturer Narrative
The reported event was inconclusive because no sample was returned.A potential root cause for this failure mode could be ''under cured sac'' (example: poor shape/ baggy sac-nonconcentric balloon).The lot number was unknown; therefore, the device history record could not be reviewed.Unable to review the labelling due to unknown product code.Although the product family was unknown, the foley catheter ifus were found to be adequate based on past reviews.The device was not returned.
 
Event Description
It was reported that the customer had a leakage issue with the foley catheter.Stated that the catheter inflated to one side and was causing this issue.And their doctor thought it was bladder spasm and tried over and under inflating the catheter but did not solve this.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LATEX FOLEY CATHETER
Type of Device
UNKNOWN FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
juan velez
8195 industrial blvd
covington 30014
7707846100
MDR Report Key15370677
MDR Text Key305176150
Report Number1018233-2022-06962
Device Sequence Number1
Product Code EZC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K910846
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/31/2022
Initial Date FDA Received09/07/2022
Supplement Dates Manufacturer Received09/26/2022
Supplement Dates FDA Received09/28/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-