This device is not distributed in us so that 510k# is blank.We checked the returned unit and confirmed that the operation channel blocked.Based on the result, we concluded that it was caused due to the insufficient reprocessing at the facility on the operation channel.In addition, we confirmed that the light guide cable buckled, the biopsy inlet t-piece clogged, and the insertion flexible tube (ift) spiral closer condition; however, they are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0588(channel)"" and/or the risk analysis results, it was evaluated to submit mdr.
|