MERIT MEDICAL SINGAPORE DTX SAFEDRAW KITS.; BP-MONITORING SET Back to Search Results
Catalog Number 688360/JPB
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/29/2022
Event Type  malfunction  
Event Description
The account alleges that when pressurizing the saline bag for a hemodynamic pressure monitoring procedure, leakage occurred from the a-line tubing connector.The connector was damaged.The device was replaced.No patient injury to report.
 
Manufacturer Narrative
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
 
Manufacturer Narrative
The suspect device was returned for evaluation.The complaint was confirmed and the root cause is attributed to the manufacturing process.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DTX SAFEDRAW KITS.
Type of Device
BP-MONITORING SET
Manufacturer (Section D)
MERIT MEDICAL SINGAPORE
198 yishun ave 7
singapore, 76892 6
SN  768926
Manufacturer (Section G)
MERIT MEDICAL SINGAPORE
198 yishun ave 7
singapore, 76892 6
SN   768926
Manufacturer Contact
bryson heaton bsn,rn.
1600 merit parkway
south jordan, UT 84095
MDR Report Key15380173
MDR Text Key299971038
Report Number8020616-2022-00063
Device Sequence Number1
Product Code DSK
UDI-Device Identifier04589731124408
UDI-Public4589731124408
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/25/2022
Device Catalogue Number688360/JPB
Device Lot NumberC2387287
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/30/2022
Initial Date FDA Received09/08/2022
Supplement Dates Manufacturer Received09/27/2022
Supplement Dates FDA Received10/06/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/25/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-