• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL HD ARTHROSCOPE/SINUSCOPE 4.0MM X 30 DEG X 167MM (MITEK LOCK); RIGID ENDOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL HD ARTHROSCOPE/SINUSCOPE 4.0MM X 30 DEG X 167MM (MITEK LOCK); RIGID ENDOSCOPE Back to Search Results
Model Number 242018
Device Problems Display or Visual Feedback Problem (1184); Use of Device Problem (1670); Device Contaminated During Manufacture or Shipping (2969)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/20/2022
Event Type  malfunction  
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).Initial reporter is a j&j sales representative.The device manufacture date is currently unavailable.Investigation summary: the complaint device was received at the manufacturing site and evaluated.Per service reports, this complaint can be confirmed.During the service evaluation the following defects were identified: outer tube damaged, distal tip damaged.Distal tip has deposits.Image cloudy/blurred.The defective parts replaced to resolve the issues.After repair, the device was found to be working according to the specifications.As part of depuy mitek¿s quality process all devices are manufactured, inspected, and released to approved specifications.The faulty parts was identified as the root cause for the device failure during the service evaluation.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Device history lot: manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.Device history batch: null.Device history review: null.
 
Event Description
It was reported by the sales rep that the hd arthroscope/sinuscope 4.0mm x 30 deg x 167mm (mitek lock) was physically damaged.During in-house engineering evaluation, it was determined that the outer tube and distal tip were damaged, the distal tip had deposits and the image was cloudy/blurred on the device.There was no procedure nor patient involvement reported.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HD ARTHROSCOPE/SINUSCOPE 4.0MM X 30 DEG X 167MM (MITEK LOCK)
Type of Device
RIGID ENDOSCOPE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle CH-24 00
SZ  CH-2400
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle CH-24 00
SZ   CH-2400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key15381586
MDR Text Key305890446
Report Number1221934-2022-02749
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705029150
UDI-Public10886705029150
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K080560
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number242018
Device Catalogue Number242018
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/10/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/23/2022
Initial Date FDA Received09/08/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-