Additional manufacturer narrative: the aquabeam scope was not returned for investigation.The user facility confirmed that the scope was accidently discarded.A review of the device history record (dhr) for aquabeam robotic system/serial number (b)(4) and the aquabeam scope/ lot number 72007/s0003 was performed, which confirmed that there were no nonconformances, failures, discrepancies, or missed steps during the manufacturing process that could be related to the reported event.The review indicated that the system and its associated component met all design and manufacturing specifications when released for distribution.The aquabeam robotic system user manual, um0101-00 rev.F, and , state the following: 11.2.5 sterile: aquabeam handpiece and aquabeam scope setup states: hold the distal end of the scope tube tip approximately 1 inch (2.54cm) from the fully proximal position and continue advancing the aquabeam scope forward until it is properly engaged with the aquabeam handpiece and then rotate the proximal key alignment adapter so that the dimple on the proximal key alignment adapter is facing up.An audible click should be heard when the aquabeam scope is securely engaged with the aquabeam handpiece.Note: do not use excessive force to advance the aquabeam scope.If resistance is felt, gently rotate the aquabeam scope clockwise and counterclockwise while simultaneously applying forward pressure.The aquabeam robotic system instruction for use, ifu0101-00 rev.E, state the following: 5.1.Precautions: general: if excessive resistance is encountered during aquabeam scope positioning, reposition the aquabeam handpiece to minimize tenting the prostate in order to reduce the chance of damaging the aquabeam scope.Based on additional information, the scope was accidentally thrown out.The root cause of the reported event could not be determined due to the inability to investigate the device.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
|