Model Number AB2000
Device Problem Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/10/2022
Event Type  malfunction  
Event Description
A male patient underwent an aquablation procedure for symptomatic benign prostatic hyperplasia (bph).Procept biorobotics corporation (procept) was made aware that during the aquablation procedure the aquabeam scope bent and could not be inserted into the aquabeam handpiece and had to be replaced with a new unit.The reported event resulted in a procedural delay of over 20 minutes.There were no adverse health consequences to the patient due to this event.
 
Manufacturer Narrative
Additional manufacturer narrative: the aquabeam scope was not returned for investigation.The user facility confirmed that the scope was accidently discarded.A review of the device history record (dhr) for aquabeam robotic system/serial number (b)(4) and the aquabeam scope/ lot number 72007/s0003 was performed, which confirmed that there were no nonconformances, failures, discrepancies, or missed steps during the manufacturing process that could be related to the reported event.The review indicated that the system and its associated component met all design and manufacturing specifications when released for distribution.The aquabeam robotic system user manual, um0101-00 rev.F, and , state the following: 11.2.5 sterile: aquabeam handpiece and aquabeam scope setup states: hold the distal end of the scope tube tip approximately 1 inch (2.54cm) from the fully proximal position and continue advancing the aquabeam scope forward until it is properly engaged with the aquabeam handpiece and then rotate the proximal key alignment adapter so that the dimple on the proximal key alignment adapter is facing up.An audible click should be heard when the aquabeam scope is securely engaged with the aquabeam handpiece.Note: do not use excessive force to advance the aquabeam scope.If resistance is felt, gently rotate the aquabeam scope clockwise and counterclockwise while simultaneously applying forward pressure.The aquabeam robotic system instruction for use, ifu0101-00 rev.E, state the following: 5.1.Precautions: general: if excessive resistance is encountered during aquabeam scope positioning, reposition the aquabeam handpiece to minimize tenting the prostate in order to reduce the chance of damaging the aquabeam scope.Based on additional information, the scope was accidentally thrown out.The root cause of the reported event could not be determined due to the inability to investigate the device.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AQUABEAM ROBOTIC SYSTEM
Type of Device
FLUID JET REMOVAL SYSTEM
Manufacturer (Section D)
PROCEPT BIOROBOTICS CORPORATION
900 island drive
suite 101
redwood city CA 94065 1494
Manufacturer (Section G)
PROCEPT BIOROBOTICS CORPORATION
900 island drive
suite 101
redwood city CA 94065 1494
Manufacturer Contact
doria esquivel
900 island drive
suite 101
redwood city, CA 94065-1494
MDR Report Key15392168
MDR Text Key305411987
Report Number3012977056-2022-00113
Device Sequence Number1
Product Code PZP
UDI-Device IdentifierB614AB20001
UDI-Public+B614AB20001/16D20191219S
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
DEN170024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 09/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAB2000
Device Catalogue NumberAB2000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/10/2022
Initial Date FDA Received09/09/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured12/19/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexMale
-
-