DEPUY IRELAND - 9616671 SMARTMIX CEMVAC; BONE CEMENT : CEMENT DELIVERY DEVICES Back to Search Results
Model Number 83-1615
Device Problem Chemical Problem (2893)
Patient Problem Insufficient Information (4580)
Event Date 12/22/2021
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on (b)(6) 2021, the patient underwent the total hip arthroplasty for the femur with the cement, the syringe, and the other company¿s cement stem.The surgeon planned to put the three cements in the syringe and agitated them.The surgeon prepared as per the procedure manual and put the cements in the syringe.However, the cement powder was stuck in the syringe nozzle, and the surgeon could not put the one pack of the cement in the syringe.Therefore, the surgeon used the other company¿s cement.Surgical delay of 30 minutes occurred.No adverse affects on the patient.
 
Manufacturer Narrative
Product complaint #(b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: a manufacturing record evaluation (nc search) was performed for the finished device 831615, lot - 21010243, and no non-conformances / manufacturing irregularities were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMARTMIX CEMVAC
Type of Device
BONE CEMENT : CEMENT DELIVERY DEVICES
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY CMW - 9610921
cornford rd
blackpool FY4 4 QQ
UK   FY4 4QQ
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key15397569
MDR Text Key304635577
Report Number1818910-2022-17571
Device Sequence Number1
Product Code KIH
UDI-Device Identifier10603295202493
UDI-Public10603295202493
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number83-1615
Device Catalogue Number831615
Device Lot Number21010243
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/18/2022
Initial Date FDA Received09/10/2022
Supplement Dates Manufacturer Received09/12/2022
Supplement Dates FDA Received09/12/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/27/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-