Model Number 0672 |
Device Problems
Signal Artifact/Noise (1036); Failure to Capture (1081); Over-Sensing (1438); Failure to Sense (1559); Low impedance (2285)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/17/2022 |
Event Type
Injury
|
Event Description
|
It was reported that the patient experienced syncopal events while at home.Review of the implantable cardioverter defibrillator (icd) data revealed that the patient was in an ongoing atrial arrhythmia.There were no alerts and the icd was operating as programmed.Further review, however, indicated that there was signal noise on the right ventricular (rv) channel and that the rv pace impedance had sporadic decreases to below 200 ohms.The noise did not appear to be sensed.Technical services recommended further troubleshooting.The rv lead remains in service and no adverse patient effects were reported.
|
|
Event Description
|
It was reported that the patient experienced syncopal events while at home.Review of the implantable cardioverter defibrillator (icd) data revealed that the patient was in an ongoing atrial arrhythmia.There were no alerts and the icd was operating as programmed.Further review, however, indicated that there was signal noise on the right ventricular (rv) channel and that the rv pace impedance had sporadic decreases to below 200 ohms.The noise did not appear to be sensed.Technical services recommended further troubleshooting.The rv lead remains in service and no adverse patient effects were reported.It was additionally reported that the icd recorded a number of ventricular high-rate episodes that appeared to be oversensing of the rv lead noise.Review of the device events indicated the rv noise had been present for some time.Provocative maneuvers/isometrics were not performed as the patient was currently in a high-dependency bed after a possible stroke.The rv lead remains in service.
|
|
Event Description
|
It was reported that the patient experienced syncopal events while at home.Review of the implantable cardioverter defibrillator (icd) data revealed that the patient was in an ongoing atrial arrhythmia.There were no alerts and the icd was operating as programmed.Further review, however, indicated that there was signal noise on the right ventricular (rv) channel and that the rv pace impedance had sporadic decreases to below 200 ohms.The noise did not appear to be sensed.Technical services recommended further troubleshooting.The rv lead remains in service and no adverse patient effects were reported.It was additionally reported that the icd recorded a number of ventricular high-rate episodes that appeared to be oversensing of the rv lead noise.Review of the device events indicated the rv noise had been present for some time.Provocative maneuvers/isometrics were not performed as the patient was currently in a high-dependency bed after a possible stroke.The rv lead remains in service.It was additionally reported that the patient was recently seen for follow-up.Upon interrogation of the icd, it was observed that the rv lead was neither sensing the intrinsic rhythm nor capturing and the pace impedance was less than 200 ohms.Rv lead noise was still being sensed.Leading to storage of tachy episodes.No inappropriate therapy had been delivered.The patient was not pacing dependent and thus was not symptomatic.The therapy was turned off and the patient was admitted to the hospital for urgent lead replacement.The lead was capped and abandoned.A new lead was successfully implanted and no additional adverse patient effects were reported.
|
|
Event Description
|
It was reported that the patient experienced syncopal events while at home.Review of the implantable cardioverter defibrillator (icd) data revealed that the patient was in an ongoing atrial arrhythmia.There were no alerts and the icd was operating as programmed.Further review, however, indicated that there was signal noise on the right ventricular (rv) channel and that the rv pace impedance had sporadic decreases to below 200 ohms.The noise did not appear to be sensed.Technical services recommended further troubleshooting.The rv lead remains in service and no adverse patient effects were reported.It was additionally reported that the icd recorded a number of ventricular high-rate episodes that appeared to be oversensing of the rv lead noise.Review of the device events indicated the rv noise had been present for some time.Provocative maneuvers/isometrics were not performed as the patient was currently in a high-dependency bed after a possible stroke.The rv lead remains in service.It was additionally reported that the patient was recently seen for follow-up.Upon interrogation of the icd, it was observed that the rv lead was neither sensing the intrinsic rhythm nor capturing and the pace impedance was less than 200 ohms.Rv lead noise was still being sensed.Leading to storage of tachy episodes.No inappropriate therapy had been delivered.The patient was not pacing dependent and thus was not symptomatic.The therapy was turned off and the patient was admitted to the hospital for urgent lead replacement.The lead was capped and abandoned.A new lead was successfully implanted, and no additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Correction: please refer to additional impact code f23: unexpected medical intervention.
|
|
Search Alerts/Recalls
|