Catalog Number C45101314J
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The sample was visually inspected and no discrepancies detected.The circuit did not pass the leak test.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.No information is provided to date.Operator of device is unknown.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Event Description
It was reported that during a pre-use check, the customer noticed air was leaking from the anesthesia circuit.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX GENERAL ANESTHESIA CIRCUITS
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
7-15-26 fukushima,
minneapolis, MN 55442
MDR Report Key15410589
MDR Text Key305730312
Report Number3012307300-2022-18928
Device Sequence Number1
Product Code BTO
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberC45101314J
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/07/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/08/2022
Initial Date FDA Received09/13/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-