SMITH & NEPHEW ORTHOPAEDICS AG BICON-PLUS SHELL; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, UNCEMENTED Back to Search Results
Catalog Number UNKNOWN
Device Problem Malposition of Device (2616)
Patient Problem Joint Dislocation (2374)
Event Date 06/11/2002
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference: (b)(4).Korovessis, p., petsinis, g., repanti, m., papazisis, z., iliopoulos, p., & soucacos, p.N.(2002).Short-term results with the zweymueller-sl metal-on-metal total hip arthroplasty.European journal of orthopaedic surgery & traumatology, 12(2), 81-89.
 
Event Description
It was reported that, on the literature review "short-term results with the zweymueller-sl metal-on-metal total hip arthroplasty", a total of 266 patients underwent an index surgery where a bicon plus shell and a sl-plus stem were implanted as part of a total hip replacement procedure.One (1) patient experienced a dislocation due to insertion of the screw socket in an increased valgus position.Initially, this was treated with an anti-rotation leg plaster, but, due to recurrent dislocations, a revision surgery was performed.Further information is unknown.
 
Manufacturer Narrative
H3, h6: in the study of korovessis, p., et.Al.[1], a total of 266 patients underwent an index surgery where a bicon plus shell and a sl-plus stem were implanted as part of a total hip replacement procedure.One (1) patient experienced a dislocation due to insertion of the screw socket in an increased valgus position.Initially, this was treated with an anti-rotation leg plaster, but, due to recurrent dislocations, a revision surgery was performed.The complaint device, used in treatment, was not received for investigation and a visual inspection could not be performed.The exact part and batch number of the device is not known.Therefore, the batch record review could not be performed and it is not possible to investigate whether the reported devices met manufacturing specifications upon release for distribution.A complaint history review was performed.The occurrence of the reported failure mode is within its expected risk level as per risk management.Furthermore, the ifu indicates several possible adverse effects resulting from a hip arthroplasty.A review of the risk management documentation verifies the failure mode and severity of the reported issue.Due to insufficient information it is not possible to perform a review of past corrective actions.For the medical investigation, no relevant supporting clinical information could be provided.Therefore, a thorough clinical assessment cannot be performed at this time.The patient's current condition is unknown and the patient impact beyond the reported events could not be determined.Should any additional clinical information be provided, this complaint will be re-evaluated.Based on the performed investigations, a relationship between the reported devices and the reported incidents cannot be excluded, but the root cause of the problem stays undetermined due to insufficient information.A relationship between the reported event and the device cannot be confirmed.Due to insufficient information it is not possible to speculate about factors which could have contributed to the reported event.To date, no further actions will be taken.Smith and nephew will monitor the devices for further similar issues.[1]: korovessis, p., petsinis, g., repanti, m., papazisis, z., iliopoulos, p., & soucacos, p.N.(2002).Short-term results with the zweymueller-sl metal-on-metal total hip arthroplasty.Résultats précoces avec la prothèse de hanche métal-métal zweymueller-sl.European journal of orthopaedic surgery & traumatology : orthopedie traumatologie, 12(2), 81¿89.Https://doi.Org/10.1007/s00590-002-0025-6 internal complaint reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BICON-PLUS SHELL
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ  CH-5000
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ   CH-5000
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15417683
MDR Text Key299854994
Report Number9613369-2022-00495
Device Sequence Number1
Product Code LWJ
Combination Product (y/n)N
Reporter Country CodeGR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Literature,Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/23/2022
Initial Date FDA Received09/14/2022
Supplement Dates Manufacturer Received10/06/2022
Supplement Dates FDA Received10/18/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNKN. SL-PLUS STEM
Patient Outcome(s) Required Intervention; Hospitalization;
-
-