Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/13/2019
Event Type  malfunction  
Event Description
It was reported that during pre-test, the cuff was inflated with sterile water and was found to be leaking.No patient injury was reported.
 
Manufacturer Narrative
Model#: udi information is unknown.Premarket (510k) number is unknown.No lot number was provided; therefore, device history record review could not be completed.A product sample was received for evaluation.Visual and functional testing were performed.A hole was found in the cuff.Unable to determine the root cause of the reported issue.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA TUBES NEO/PED FLEXTEND PLUS
Type of Device
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15418172
MDR Text Key305893334
Report Number3012307300-2022-19135
Device Sequence Number1
Product Code JOH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/25/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/08/2022
Initial Date FDA Received09/14/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-